We report a case of acute arterial ischemia and deep venous thrombosis due to compression of the external iliac vein and artery by the reservoir of a three-piece inflatable penile prosthesis. Deho' F, Henry GD, Marone EM, Sacca' A, Chiesa R, Rigatti P, and Montorsi F. Severe vascular complication after implantation of a three-piece inflatable penile prosthesis. © 2008 International Society for Sexual Medicine.
Severe vascular complication after implantation of a three-piece inflatable penile prosthesis
MARONE, ENRICO MARIA;
2008-01-01
Abstract
We report a case of acute arterial ischemia and deep venous thrombosis due to compression of the external iliac vein and artery by the reservoir of a three-piece inflatable penile prosthesis. Deho' F, Henry GD, Marone EM, Sacca' A, Chiesa R, Rigatti P, and Montorsi F. Severe vascular complication after implantation of a three-piece inflatable penile prosthesis. © 2008 International Society for Sexual Medicine.File in questo prodotto:
Non ci sono file associati a questo prodotto.