Temporary anchorage devices (TADs) have been introduced into orthodontic clinical practice in order to allow tooth movements while avoiding strain on adjacent teeth. Miniscrews are available in the market with different diameters and materials. Accordingly, the purpose of the present report was to measure and compare the forces to bend and fracture different mini implants. Ti-6Al-4V titanium and stainless steel TADs of different manufacturers (Spider ScrewHDC; Mini Implants⁻Leone; Benefit⁻Orteam; Storm⁻Kristal) were evaluated. Two different diameters (1.5 mm and 2.0 mm) were tested. The sample included 10 unused specimens for each group, blocked in an Instron Universal Testing Machine, and a shear load was applied at the neck of the miniscrew. The force to bend the miniscrew was measured at 0.1 mm and 0.2 mm deflections. Also, the maximum force before screw fracture was recorded. Data were submitted for statistical analysis. Results showed significantly higher forces for 2.0 mm than 1.5 mm screws, both at 0.1 mm and 0.2 mm deflections and at maximum load. Moreover, no significant differences were reported between titanium and stainless steel miniscrews of equal diameters.

Reliability of Orthodontic Miniscrews: Bending and Maximum Load of Different Ti-6Al-4V Titanium and Stainless Steel Temporary Anchorage Devices (TADs)

Scribante, Andrea;Bernardinelli, Luisa;ALCOZER, ROBERTO;Gandini, Paola;Sfondrini, Maria Francesca
2018-01-01

Abstract

Temporary anchorage devices (TADs) have been introduced into orthodontic clinical practice in order to allow tooth movements while avoiding strain on adjacent teeth. Miniscrews are available in the market with different diameters and materials. Accordingly, the purpose of the present report was to measure and compare the forces to bend and fracture different mini implants. Ti-6Al-4V titanium and stainless steel TADs of different manufacturers (Spider ScrewHDC; Mini Implants⁻Leone; Benefit⁻Orteam; Storm⁻Kristal) were evaluated. Two different diameters (1.5 mm and 2.0 mm) were tested. The sample included 10 unused specimens for each group, blocked in an Instron Universal Testing Machine, and a shear load was applied at the neck of the miniscrew. The force to bend the miniscrew was measured at 0.1 mm and 0.2 mm deflections. Also, the maximum force before screw fracture was recorded. Data were submitted for statistical analysis. Results showed significantly higher forces for 2.0 mm than 1.5 mm screws, both at 0.1 mm and 0.2 mm deflections and at maximum load. Moreover, no significant differences were reported between titanium and stainless steel miniscrews of equal diameters.
2018
Medical Research, Organs & Systems includes resources dealing with the normal and disease states of single organs, tissues, or single physiological systems, exclusive of the heart, vascular and immune systems. Systems covered here include hepatology, pulmonary function/physiology, gastroenterology, otolaryngology, respiratory system, andrology, gynecology and reproduction, dermatology, and dentistry/odontology. Resources dealing with general physiology, classes of disease that immediately affect many or all body systems, and medical research focused on specific types of medical intervention are excluded.
Comitato scientifico
Inglese
Internazionale
ELETTRONICO
11
7
1138
10
anchorage; bend; dentistry; fracture; implant; load; miniscrew; orthodontics; shear
7
info:eu-repo/semantics/article
262
Scribante, Andrea; Montasser, Mona A; Radwan, Eman Saad; Bernardinelli, Luisa; Alcozer, Roberto; Gandini, Paola; Sfondrini, Maria Francesca
1 Contributo su Rivista::1.1 Articolo in rivista
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11571/1222566
Citazioni
  • ???jsp.display-item.citation.pmc??? 14
  • Scopus 27
  • ???jsp.display-item.citation.isi??? 22
social impact