We have built a number of test arrays made of 16 SiPMs of 3 × 3 mm2 with a mixed series–parallel configuration and common bias and readout. With this technique it is possible to increase the total active area keeping low the bias voltage and readout channels. To further increase the total area we connected few arrays together. Tests were performed in terms of pulse amplitude, charge and timing features.

Characterization of SiPM arrays with common bias and common readout for applications in liquid argon

Cervi T.;Menegolli A.;
2019-01-01

Abstract

We have built a number of test arrays made of 16 SiPMs of 3 × 3 mm2 with a mixed series–parallel configuration and common bias and readout. With this technique it is possible to increase the total active area keeping low the bias voltage and readout channels. To further increase the total area we connected few arrays together. Tests were performed in terms of pulse amplitude, charge and timing features.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11571/1394314
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact