Magnetic resonance imaging (MRI) is a widely used diagnostic technique. Patients wearing orthodontic devices are often requested to remove their appliances before an MRI exam, even when the exam involves anatomical areas far from the head, in order to prevent the heating and detachment of the appliances. The present report aims to evaluate changes in temperature and adhesive forces of molar bands after MRI at two different strength outputs. Sixty stainless steel molar bands were bonded on permanent human upper molars using two different cements: Unitek Multi-Cure Glass Ionomer Band Cement (3M Unitek, Monrovia, CA, USA) and Transbond Plus Light Cure Band Adhesive (3M Unitek). Appliances were subjected to MRI with two different strengths (1.5 Tesla and 3 Tesla). Tubes and band temperature was measured before and after MRI. Subsequently, the shear bond strength (SBS) test was calculated. Data underwent statistical analysis (p < 0.05). After MRI, molar bands exhibited significant heating, even though not clinically relevant, with a temperature increase ranging between 0.48 degrees C and 1.25 degrees C (p < 0.05). SBS did not show significant differences (p > 0.05). The present study suggests that, under MRI, the molar bands tested are safe; therefore, their removal could be not recommended for non-head and neck MRI exams. Removal would be necessary just in artifact risk areas.

Effect of Magnetic Resonance Imaging at 1.5 T and 3 T on Temperature and Bond Strength of Orthodontic Bands with Welded Tubes: An In Vitro Study

Sfondrini, Maria Francesca;Gallo, Simone;Pascadopoli, Maurizio;Rizzi, Cinzia;Gorone, Maria Sole Prevedoni;Preda, Lorenzo;Gandini, Paola;Scribante, Andrea
2023-01-01

Abstract

Magnetic resonance imaging (MRI) is a widely used diagnostic technique. Patients wearing orthodontic devices are often requested to remove their appliances before an MRI exam, even when the exam involves anatomical areas far from the head, in order to prevent the heating and detachment of the appliances. The present report aims to evaluate changes in temperature and adhesive forces of molar bands after MRI at two different strength outputs. Sixty stainless steel molar bands were bonded on permanent human upper molars using two different cements: Unitek Multi-Cure Glass Ionomer Band Cement (3M Unitek, Monrovia, CA, USA) and Transbond Plus Light Cure Band Adhesive (3M Unitek). Appliances were subjected to MRI with two different strengths (1.5 Tesla and 3 Tesla). Tubes and band temperature was measured before and after MRI. Subsequently, the shear bond strength (SBS) test was calculated. Data underwent statistical analysis (p < 0.05). After MRI, molar bands exhibited significant heating, even though not clinically relevant, with a temperature increase ranging between 0.48 degrees C and 1.25 degrees C (p < 0.05). SBS did not show significant differences (p > 0.05). The present study suggests that, under MRI, the molar bands tested are safe; therefore, their removal could be not recommended for non-head and neck MRI exams. Removal would be necessary just in artifact risk areas.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11571/1472255
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact