This study measures the resistance to gas flow of different double-lumen bronchial tubes used for separate ventilation of each lung in critically ill patients. Different-sized Robertshaw and Carlens' tubes were studied, as well as a new device that consisted of a cuffed bronchial catheter introduced through a standard tracheal tube. The pressure-flow relationship was curvilinear in all cases. Robertshaw and Carlens' tubes were generally found to have bronchial and tracheal channels with almost similar resistance and offered total resistances similar to those of 7.0-8.0 mm internal diameter tracheal tubes. The new device had a higher resistance than others of equal external diameter. This must be taken into account when its use is recommended for prolonged respiratory support.

Measurement of the flow-resistive properties of double-lumen bronchial tubes in vitro.

BRASCHI, ANTONIO
1989-01-01

Abstract

This study measures the resistance to gas flow of different double-lumen bronchial tubes used for separate ventilation of each lung in critically ill patients. Different-sized Robertshaw and Carlens' tubes were studied, as well as a new device that consisted of a cuffed bronchial catheter introduced through a standard tracheal tube. The pressure-flow relationship was curvilinear in all cases. Robertshaw and Carlens' tubes were generally found to have bronchial and tracheal channels with almost similar resistance and offered total resistances similar to those of 7.0-8.0 mm internal diameter tracheal tubes. The new device had a higher resistance than others of equal external diameter. This must be taken into account when its use is recommended for prolonged respiratory support.
1989
The Anesthesia & Intensive Care category covers resources concerned with all aspects of anesthesia including the delivery and administration of anesthetics, emergency medicine, critical care medicine, traumatology, burns, injury, and resuscitation.
Sì, ma tipo non specificato
Inglese
Internazionale
STAMPA
44
335
340
Airway Resistance, Bronchi, Humans, Mathematics, Pressure, Respiration; Artificial; instrumentation/methods, Rheology
5
info:eu-repo/semantics/article
262
M., Chiaranda; A., Rossi; G., Manani; O., Pinamonti; Braschi, Antonio
1 Contributo su Rivista::1.1 Articolo in rivista
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11571/444319
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 6
social impact